Model 90M9
mAs Sensor
for the 9096 and 4085/6
SPECIFICATIONS:
• Full-scale range : 2000 mA or 9999 mAs
• mA accuracy: 0.2% of reading at DC plus resolution of +/-0.015 mA or 3-4 digits
• mAs accuracy (1-s pulse): 5 чAs or 0.2% of reading; resolution of 0.001 mAs.
• Input impedance: Fixed, resistive, 1.7 - 2.7 ohms.
1 ohm sense resistor plus fuse and wiring.
• Input protection: 2-A fuse.
Over-voltage protection using 22-J varistor before the fuse and 17-V Zener diodes after the fuse.
• Absolute value of bi-polar 720 Hz waveform accurate to 2.5%.
• Bandwidth: 2.33 kHz, -3 dB.
• ON/OFF controlled from 9095/6 or 4085/6 control unit.
(ON only when a kV function is active.)
• Power supplied by replaceable 9V battery |