RADCAL HASTA DOZ KALİBRATÖRÜ
PDC
|
|
Assessment of an Instrument for Calibrating Dose Area Product (DAP) Meters
Radcal Corporation (California, USA) have developed a "Patient Dose Calibrator", (PDC) designed to allow quick and easy calibration checks of installed DAP meters.
The PDC method of calibration is significantly quicker than our traditional method, and allows a full check of the DAP meter calibration over a range of field sizes and shapes and beam quaities. |
The Route To Patient Dose Control |
|
|
|
• |
The PDC solves much publicized problems relating to variation of DAP, Patient Dose system and AEC calibration. |
|
• |
Measures DAP, DAP-rate, Dose and Dose-rate. |
• |
Ideally suited to aid the setting of reference doses. |
• |
Widespread use of the PDC by equipment manufacuturers, x-ray service organizations, physicists and radiation technologists will help acheive consistent patient dose levels. |
• |
Ion chamber performance. |
• |
Optical and radiographic alignment markers. |
• |
Simple to use with optional computer control. |
|
• DAP and dose linearity with dose and field size.
• Sensitivity testing.
• DAP and dose calibration at the patient plane or at a reference distance.
|
|
• Entrance surface dose calibration for different examinations using a phantom.
• Establishing examination related reference DAP and dose levels. |
|
• Image receptor dose measurement.
• AEC linearity testing. | |
|
Technical Data |
|
Display range Dose area product Air kerma
|
(0.01... 99 999 999)µGy·m3 (0.001... 99 999 999)mGy
|
Accuracy DAP Air kerma |
± 10 % ± 10 % (15cm x 15cm field) |
Inclusive of all uncertainties (temperature, pressure, rate and area dependencies and beam quality. |
Digital Resolution Dose area product Dose area product rate Air kerma Air kerma rate |
0.01 µGy·m2 1µGy·m2/min 0.001mGy 0.1 mGy/min
|
Rated range of use Tube voltage Dose area product: (low rate range) (high rate range) Air kerma rate |
(40 ... 150) kV 1µGy·m2/min (1 ... 1·104) µGy·m2/min (2·103 ... 9 ·105) µGy·m2/min (0.2 ... 9 ·103) mGy/min (at the position of the chamber) |
Automatic Temperature and Pressure Correction |
Pressure Temperature Air humidity
|
(80.0 ... 106.0) kPa (+10 ... +40) ºC (10 ... 80) % rel. humidity (max. 20 g/m3) | |
Ionization chamber Response versus radiation quality (50kV ... 150 kV, norm. to 100kV; acc. IEC 60580)
Quality equivalent filtration (70kV) Active area Dose area product Air kerma |
± 3 %
0.6 mm Al
max. (300 x 300) mm2 min. (100 x100) mm2 |
Power supply Internal rechargeable battery pack operation time (state of charge: 100%) |
Li-ION, 2 cells > 8 h |
Serial Interface |
USB |
Protection class (acc. IEC 60529) |
IP 41 |
Weight |
2.32 kg |
Dimension |
350 mm x 410 mm x 35 mm (L x W x H) |
Note: All specifications subject to change. | | |
|
Ürüne ait PDF dosyası yok...
|